4.12. Year 12, 2000
During the 2000 cruises two developmental versions
of the Sea-Bird SBE-43 oxygen sensor were used (#43071 and #43081). These
sensors were calibrated at Sea-Bird before each cruise. In addition, two
YSI Inc. probes (#13434 and #13251) were used. Sensor #13434 had its membrane
and electrolyte replaced prior to HOT-114.
Three sets of parameters were obtained per HOT
cruise, corresponding to the casts at Stations 1, 2 and 8. Casts from
HA-8A were calibrated with bottle data obtained during that cruise.
Table 4.12 gives the means and standard deviations
for the final calibrated CTD oxygen minus water sample values. Dual sensors
were used during cruises, but only the sensor whose data were deemed more
reliable is included.
Table 4.12a: CTD-Bottle dissolved oxygen per cruise
at Station Kahe (µmol kg-1)
|
|
|
0 to 1500 dbar |
|
Cruise |
Sensor |
Mean |
SD |
| HOT-111 |
43071 |
0.01 |
1.30 |
| HOT-112 |
13434 |
0.01 |
1.20 |
| HOT-113 |
43071 |
0.02 |
1.54 |
| HOT-114 |
13434 |
0.02 |
0.97 |
| HOT-115 |
43081 |
0.00 |
0.62 |
| HOT-116 |
43081 |
0.00 |
1.32 |
| HOT-117 |
43081 |
0.00 |
0.96 |
| HOT-118 |
43081 |
0.00 |
1.05 |
| HOT-119 |
43081 |
0.00 |
0.28 |
| HOT-120 |
13434 |
0.01 |
1.13 |
|
HOT-121 |
13434 |
0.01 |
0.70 |
Table 4.12b: CTD-Bottle dissolved oxygen per cruise at Station
ALOHA (µmol kg-1)
|
|
|
0 to 4800 dbar |
500 to 4800 dbar |
|
Cruise |
Sensor |
Mean |
SD |
Mean |
SD |
| HOT-111 |
43071 |
0.09 |
1.18 |
0.13 |
1.07 |
| HOT-112 |
13434 |
0.07 |
0.98 |
0.15 |
0.81 |
| HOT-113 |
43071 |
0.13 |
1.25 |
0.18 |
1.03 |
| HOT-114 |
43081 |
0.01 |
0.80 |
-0.02 |
0.77 |
| HOT-115 |
43081 |
0.09 |
0.83 |
0.11 |
0.75 |
| HOT-116 |
43081 |
0.01 |
0.81 |
-0.02 |
0.69 |
| HOT-117 |
43081 |
0.09 |
0.85 |
0.10 |
0.76 |
| HOT-118 |
43081 |
0.01 |
0.83 |
-0.01 |
0.78 |
| HOT-119 |
43081 |
0.02 |
0.95 |
0.00 |
0.86 |
| HOT-120 |
13434 |
0.04 |
1.13 |
0.01 |
1.21 |
|
HOT-121 |
13434 |
0.04 |
1.03 |
0.02 |
1.09 |
Table 4.12: CTD-Bottle dissolved oxygen at Station HALE-ALOHA (µmol
kg-1)
|
|
|
0-1500 dbar |
|
Cruise |
Sensor |
Mean |
SD |
| HA-8A |
43081 |
0.00 |
0.95 |
| HOT-117 |
43081 |
-0.06 |
0.96 |
| HOT-118 |
43081 |
0.00 |
0.27 |
| HOT-119 |
43081 |
0.00 |
0.83 |
| HOT-120 |
13434 |
0.00 |
0.58 |
|
HOT-121 |
13434 |
0.00 |
0.49 |
Figure 4.13 shows the CTD oxygen data quality
for every 2-dbar bin in each cast at Station ALOHA for cruises 111 through
121.

Figure 4.13. CTD oxygen data quality per cast (vertical bars) and
per 2-dbar bin at Station ALOHA during HOT cruises 111 through 120. The
tick marks along the x-axis separate different casts. Casts from the same
cruise are separated by vertical black lines.
Back to Table of Contents
 |